Hi-Pressure Super Alloy Tube with flange KIT, 55 ODx20 IDx580 L (mm) , 1100oC Max. - Tube-HP-55-20-580

Hi-Pressure Super Alloy Tube with flange KIT, 55 ODx20 IDx580 L (mm) , 1100oC Max. - Tube-HP-55-20-580

RM0.00 MYR
Sale price  RM0.00 MYR Regular price 
Skip to product information
Hi-Pressure Super Alloy Tube with flange KIT, 55 ODx20 IDx580 L (mm) , 1100oC Max. - Tube-HP-55-20-580

Hi-Pressure Super Alloy Tube with flange KIT, 55 ODx20 IDx580 L (mm) , 1100oC Max. - Tube-HP-55-20-580

SKU: tube-hp-55-20-580

Price: RFQ

Description

Ni-based superalloy tube vessel (50mm O.D x 20 mm I.D x 580 mm L). It can be used under ultra-highly oxidizing and inert gas atmospheres up to 1100°C to process special compounds. It is consumable part of OTF-1200X-HP-55 which is a compact HIP (hot isostatic pressure) furnace for researching the new generation of materials at low cost. Specification: Outside Diameter 50 mm Inside Diameter 20 mm Tube Length 580 mm Chamber Length 550 mm Please click picture below to see detail of pressure vessel Max. working temperature without load 1100 °C Working Pressure & Temperature Maximum pressure inside the tube (sf = 2) Heating temperature Max. 3045 PSI  (21 Mpa) 800°C Max. 1812 PSI  (12.5 Mpa) 900°C Max. 914 PSI   (6.3 Mpa) 1000°C Max. 609 PSI   (4.2 Mpa) 1100°C The allowable pressure is calculated based on Barlow's formula. S = yield strength. Safety factory= 2 Warning : The suitable tightening torque for turning the screws on this tube is the 40-foot pound. Click the picture below for torque wrench detail. Please set up a pressure alarm point in the digital pressure display before the operation Must install a pressure regulator on the gas cylinder to increase pressure gradually. Service-life design for 2,000 hours at 800oC. The high-pressure vessel is a consumable part, please order replacement once found deformation Standard Package Ni-Based Super Alloy Seamless Tube with Super-alloy flange Optional Optional: Adjustable High-Pressure Relief Valve 1000 - 10,000 PSI (7 - 70 Mpa) Optional: Digital Vacuum/Pressure Gauge, M14 x 1.5 from -100Kp to 30Mp Optional: Solenoid valve , pressure gauge, and controller set to Relief Pressure at the desired setting. High-pressure Relief Valve Digital Pressure Gauge                    Solenoid valve Set Attention When the pressure reaches the alarm set point (Max.15Mpa), the pressure inside the vessel will be released via the solenoid valve installed on the flange. For Max. pressure >15 Mpa, please shut off the manual needle valve nearby the solenoid valve. Considering the toughness of Ni-Based Alloy, you should tighten the screws in diagonal sequence and apply equal pressure. Do not over tighten, always use the recommended torque screwdriver.

    Outside Diameter          50 mm
    Inside Diameter          20 mm
    Tube Length                          580 mm
   Chamber Length
  • 550 mm
  • Please click picture below to see detail of pressure vessel
          
           
    Max. working temperature
    without load 
         1100 °C
     Working Pressure
             &
    Temperature
 
Maximum pressure inside the tube (sf = 2)
Heating temperature
Max. 3045 PSI  (21 Mpa) 800°C
Max. 1812 PSI  (12.5 Mpa) 900°C
Max. 914 PSI   (6.3 Mpa) 1000°C
Max. 609 PSI   (4.2 Mpa) 1100°C

The allowable pressure is calculated based on Barlow's formula. S = yield strength. Safety factory= 2
Warning
  • The suitable tightening torque for turning the screws on this tube is the 40-foot pound. Click the picture below for torque wrench detail.
  • Please set up a pressure alarm point in the digital pressure display before the operation
  • Must install a pressure regulator on the gas cylinder to increase pressure gradually.
  • Service-life design for 2,000 hours at 800oC. The high-pressure vessel is a consumable part, please order replacement once found deformation
           
 Standard Package
  • Ni-Based Super Alloy Seamless Tube with Super-alloy flange
      Optional                      
                                                                                                                 
           High-pressure  Relief Valve                         Digital Pressure Gauge                    Solenoid valve Set              
      Attention
  • When the pressure reaches the alarm set point (Max.15Mpa), the pressure inside the vessel will be released via the solenoid valve installed on the flange.                                                                                                               
  •  For Max. pressure >15 Mpa, please shut off the manual needle valve nearby the solenoid valve.
  • Considering the toughness of Ni-Based Alloy, you should tighten the screws in diagonal sequence and apply equal pressure.
  • Do not over tighten, always use the recommended torque screwdriver.
,
Maximum pressure inside the tube (sf = 2)
Heating temperature
Max. 3045 PSI  (21 Mpa) 800°C
Max. 1812 PSI  (12.5 Mpa) 900°C
Max. 914 PSI   (6.3 Mpa) 1000°C
Max. 609 PSI   (4.2 Mpa) 1100°C