Description
GSL-1700X-S60HG tube furnace is designed for processing material under hydrogen gas up to 1700°C safely. The 60 mm OD processing tube is made of pure alumina, and one hydrogen gas detector is included for cutting off hydrogen gas via a built-in solenoid valve immediately upon detection of hydrogen gas leaks. The furnace can be used with all kinds of inert gasses and oxygen gas as well. The precision temperature controller provides up to 30 programmable segments for the heating profile.
SPECIFICATIONS
|
Furnace Structure
|
|
||||||
| Power |
3 KW Max. ( 20A breaker required) |
||||||
| Voltage | AC 208-240 Single Phase, 50/60 Hz | ||||||
| Max. Temperature | 1700°C ( < 1 hour ) | ||||||
| Continuous Temperature | 1600°C | ||||||
| Max. Heating Rate | <= 10°C /min (5C/min is recommended to avoid thermal-shocking the alumina tube) | ||||||
| Processing Tube |
|
||||||
| Heating Zone length | 125 mm (5") | ||||||
| Constant temp. zone | 50 mm (2") within ± 2° | ||||||
|
Temperature controller
|
|
||||||
| Temperature Accuracy | +/ - 1°C | ||||||
|
Flanges and Gas Inlet
|
One pair stainless steel vacuum flange is installed on two ends of the tube which allow high vacuum up to 10E-5 torr and 5 PSI positive pressure.
|
||||||
|
|||||||
Hydrogen Detector System![]() |
|
||||||
| Heating Element | |||||||
Laptop & software (Optional)
|
|
||||||
| Compliance |
|
||||||
Warning![]() click here to learn more about the installation of a gas regulator |
|