1250C Small Box Furnace (6x6 x7", 4.2 liter) with Programmable Controller - KSL-1200X-J

1250C Small Box Furnace (6x6 x7", 4.2 liter) with Programmable Controller - KSL-1200X-J

SKU: ksl1200xj

Price: RFQ

Specifications

SPECIFICATIONS: Structure High purity alumina fiber insulation surrounding the chamber provides maximum energy efficiency. Heavy-duty double-layer structure with a cooling fan to keep furnace exteriors safe to touch. Doble layer steel case with forced-air cooling to ensure lower safer face temperature. One venting port is built on the top of the furnace One air inlet is located at the back of the furnace. This furnace can be put into a standard glovebox Power Voltage: 110V or 208 -240V Single Phase 50/60 Hz 2.5KW (not suitable for running at 110V from wall plug) Note: A power cable is included but without a plug. Please install a plug on your own or click the link below the picture to choose a plug according to your country or your lab requirement Operating Temp Range 1250°C < 1 hour 100-1200°C continuously Stability at 1000 ° C +/- 1.0 ° C Uniformity at 1000 ° C +/- 2.0 ° C Max. Heating Rate 30°C/ min. Thermocouple K types with a standard plug: 4" Length, 3mm OD. Temperature Controller The digital temperature controller is included. Proportional–integral–derivative control (PID control) and auto-tune function 30 segments programmed with ramping, cooling, and dwelling steps Built-in over-temperature alarm and thermocouple failure alarm +/- 1 ºC temperature control accuracy Default DB9 PC communication port Temperature controller (optional) MTS02-Y Temperature control software kit (for YD518P series controllers) + 15" laptop package is available at the options bar. Click for the detailed information about the MTS02-Y software kit. Eurotherm temperature controller with +/- 0.1oC accuracy is available at the extra cost Compliance CE Certified NRTL (UL61010) or CSA certification is available upon request at extra cost. Application Notes For the longer service life of the furnace, we recommend the heating rate not to exceed 10 °C /min. The cooling rate should not exceed 10 °C /min either. Small cracks may appear on the surface of the refractory ceramics over extended use. This is a normal occurrence and the cracks may be repaired with the alumina coating The heating module is easy to replace. Click the pic left to order a spare Heating module for replacement If firing material release vapor or smoking, please put the furnace into a fume hood as left 2 pictures. Note Toxic, corrosive, or explosive gases are not allowed for use with this furnace. You may consider the furnace with a separate controller to use in a glovebox or remote distance (click Pic. 1 to order ) If you need to open a hole in the top of the furnace ( Pic. 2), please let us know. The furnace can be modified into a hybrid tube furnace (Pic. 3) Pic. 1 Pic. 2 Pic. 3

Structure
 
  • High purity alumina fiber insulation surrounding the chamber provides maximum energy efficiency. 
  • Heavy-duty double-layer structure with a cooling fan to keep furnace exteriors safe to touch. 
  • Doble layer steel case with forced-air cooling to ensure lower safer face temperature.
  • One venting port is built on the top of the furnace
  • One air inlet is located at the back of the furnace.
  • This furnace can be put into a standard glovebox
Power
Voltage: 110V or 208 -240V Single Phase 50/60 Hz
2.5KW (not suitable for running at 110V from wall plug)
Note: A power cable is included but without a plug. Please install a plug on your own or click the link below the picture to choose a plug according to your country or your lab requirement
Operating Temp Range
  • 1250°C < 1 hour
  • 100-1200°C continuously
Stability at 1000 °C +/- 1.0 °C
Uniformity at 1000 °C +/- 2.0 °C
Max. Heating Rate 30°C/ min.
Thermocouple K types with a standard plug: 4" Length, 3mm OD.
Temperature Controller
  • The digital temperature controller is included.
  • Proportional–integral–derivative control (PID control) and auto-tune function
  • 30 segments programmed with ramping, cooling, and dwelling steps
  • Built-in over-temperature alarm and thermocouple failure alarm
  • +/- 1 ºC temperature control accuracy
  • Default DB9 PC communication port
Temperature controller (optional)
 
Compliance

  • CE Certified
  • NRTL (UL61010) or CSA certification is available upon request at extra cost.
Application Notes

  • For the longer service life of the furnace, we recommend the heating rate not to exceed 10 °C /min.
  • The cooling rate should not exceed 10 °C /min either.
  • Small cracks may appear on the surface of the refractory ceramics over extended use. This is a normal occurrence and the cracks may be repaired with the alumina coating 
  • The heating module is easy to replace. Click the pic left to order a spare Heating module for replacement
  • If firing material release vapor or smoking, please put the furnace into a fume hood as left 2 pictures. 
Note

  • Toxic, corrosive, or explosive gases are not allowed for use with this furnace. You may consider the furnace with a separate controller to use in a glovebox or remote distance (click Pic. 1 to order ) 
  • If you need to open a hole in the top of the furnace ( Pic. 2), please let us know.
  • The furnace can be modified into a hybrid tube furnace (Pic. 3)
Pic. 1Pic. 2 Pic. 3