Skip to product information
7KW Induction Heating System (80mm OD Tube, up to 1900ºC) with Temperature-Control  - SP7TC

7KW Induction Heating System (80mm OD Tube, up to 1900ºC) with Temperature-Control - SP7TC

SKU: sp7tc

Price: RFQ

Specifications

EQ-SP-7TC is a compact temperature-controlled induction heating/melting system with complete accessories, which includes a 7KW Induction Heater, 80mm O.D Quartz Tube with Flanges & Valves, Graphite Crucible & Refractory Assembly, 58L/m Recirculating Water Chiller, Vacuum Pump, and Mobile Cart. It is an ideal tool for melting, sintering new metallic alloys, and melt infiltration with a melting point < 1900ºC (under vacuum or protective gas environment). SPECIFICATIONS: Induction Heater Working Voltage: 208~240V AC, 50/ 60Hz, Single Phase Working Current: 34 Amp (60A breaker required) Max. Input Power: 7 KW Output Frequency: 30 - 80 kHz Adjustable Duty Cycle: 80% Air Cooling: Rear Fan Protection: Automatic water pressure, over-temperature, and overload protection Refractory Assembling: Complete refractory assembling made of alumina, including bottom crucible support, crucible holder, and cover Temp. Controller & Thermocouple Working Voltage: 208~240V AC, Single Phase Automatic PID tunning for temperature control. 30 segments programmable for precise controls of heating rate, cooling rate* (won't get faster than natural cooling) and dwell time. Overheating & thermocouple failure protection. +/- 5 ºC temperature accuracy. Thermocouple: C type Continuous Working Temp. Range: 1000ºC~2000ºC Max. Heating Rate: 10ºC/second (1000ºC-1200ºC) 8ºC/second (1200ºC~1500ºC) 3ºC~6ºC/second (1500ºC~1700ºC) Graphite Crucible & liners High Purity Graphite Crucible: 1.5"(OD) x 1.25"(ID) x 3.75"(Deep) Crucible is included for induction heating and metal melting. Max. Loading Capacity: Up to 1/2 of the total volume of the graphite crucible, ~35ml. Double-layer refractory liner: Graphite fiber sheet is a thermal insulator The crucible for melt infiltration is available upon request Sealing Flanges & Quartz Chamber Quartz Tube: Two-end Open Tube is included with dimensions of 80 mm(OD) x 72 mm(ID) x 365mm(L) Top Flange: Stainless steel hinged type flange with 1/4'' thermocouple feedthrough, needle valve, and barbed fitting. Please click the picture (bottom-right) to view the details. Bottom Flange: Stainless steel flange with needle valve and KF-25 vacuum port. Please click the picture (bottom-left) to view the details. KF25 right angle valve & clamps and 1 meter SS bellows are included for connecting to the vacuum pump Pressure Sensor (optional) Anti-Corrosive Capacitance Diaphragm Gauge, the measurement range is up to 3.8x10E-5 torr Water Chiller Temperature range: 5 - 30°C Water Flow Rate: 16 L/min Water Tank: 12 Liters made of stainless steel Note: It is recommended to use a specific coolant to get the best performance. Vacuum Pump One UL Certified 156 L/min Double-Stage Rotary Vane Vacuum Pump is included Vacuum level : 5x10e-2 torr by the mechanical vacuum. For better melting may choose turbopump at extra cost to reach 10e-5 torr vacuum Heavy Duty Mobile Cart Structure: Steel Frame Mobile Cart 600mm(L) x 800mm(W) surface for placing Induction Heater & Quartz Tube with Flange Compliance NRTL Certification is available upon request at extra cost, please contact our sales representatives for a price quote. Certify standards include: 1) Induction heater: UL61010-1 & UL499; 2) Vacuum Pump: UL1450; 3) Water Chiller: UL61010 & UL1995 Application Notes Click here to learn more about the installation of a gas regulator. Please click here to learn the Fundamentals of Induction Heating The Induction heating system is designed for use under vacuum and low pressure < 0.02 MPa. Attention : A pressure regulator must be installed on your input gas tank ( click picture below to review ) and have the output gas pressure set below 3 PSI at all times after opening the gas tank with the gas maintained in a continuous flowing condition. Never heat the furnace while the gas release valve is closed and a positive pressure condition exists within the furnace chamber. The pressure gauge must be used to closely monitor the chamber condition at all times during the healing process The flow rate for gases should be limited to < 200 SCCM ( or 200ml/min )

Induction Heater
  • Working Voltage: 208~240V AC, 50/ 60Hz, Single Phase
  • Working Current: 34 Amp (60A breaker required)
  • Max. Input Power: 7 KW
  • Output Frequency: 30 - 80 kHz Adjustable
  • Duty Cycle: 80%
  • Air Cooling: Rear Fan
  • Protection: Automatic water pressure, over-temperature, and overload protection
  • Refractory Assembling: Complete refractory assembling made of alumina, including bottom crucible support, crucible holder, and cover
Temp. Controller & Thermocouple
  • Working Voltage: 208~240V AC, Single Phase
  • Automatic PID tunning for temperature control.
  • 30 segments programmable for precise controls of heating rate, cooling rate* (won't get faster than natural cooling) and dwell time.
  • Overheating & thermocouple failure protection.
  • +/- 5 ºC temperature accuracy.
  • Thermocouple: C type
  • Continuous Working Temp. Range: 1000ºC~2000ºC
  • Max. Heating Rate: 
    • 10ºC/second (1000ºC-1200ºC)
    • 8ºC/second (1200ºC~1500ºC)
    • 3ºC~6ºC/second (1500ºC~1700ºC)
Graphite Crucible & liners
  • High Purity Graphite Crucible: 1.5"(OD) x 1.25"(ID) x 3.75"(Deep) Crucible is included for induction heating and metal melting.
  • Max. Loading Capacity: Up to 1/2 of the total volume of the graphite crucible, ~35ml.
  • Double-layer refractory liner: Graphite fiber sheet is a thermal insulator
  • The crucible for melt infiltration is available upon request 
Sealing Flanges & Quartz Chamber
     
  • Quartz Tube: Two-end Open Tube is included with dimensions of 80 mm(OD) x 72 mm(ID) x 365mm(L)
  • Top Flange: Stainless steel hinged type flange with 1/4'' thermocouple feedthrough, needle valve, and barbed fitting. Please click the picture (bottom-right) to view the details.
  • Bottom Flange: Stainless steel flange with needle valve and KF-25 vacuum port. Please click the picture (bottom-left) to view the details.
  • KF25 right angle valve & clamps and 1 meter SS bellows are included for connecting to the vacuum pump
               
Pressure Sensor (optional)
  • Anti-Corrosive Capacitance Diaphragm Gauge, the measurement range is up to 3.8x10E-5 torr
Water Chiller
 
  • Temperature range: 5 - 30°C
  • Water Flow Rate: 16 L/min
  • Water Tank: 12 Liters made of stainless steel
  • Note: It is recommended to use a specific coolant to get the best performance.
Vacuum Pump
 
       
Heavy Duty Mobile Cart
 
  • Structure: Steel Frame Mobile Cart
  • 600mm(L) x 800mm(W) surface for placing Induction Heater & Quartz Tube with Flange
Compliance
  • NRTL Certification is available upon request at extra cost, please contact our sales representatives for a price quote.
  • Certify standards include: 1) Induction heater: UL61010-1 & UL499; 2) Vacuum Pump: UL1450; 3) Water Chiller: UL61010 & UL1995

Application Notes

Click here to learn more about the installation of a gas regulator.

  • Please click here to learn the Fundamentals of Induction Heating
  • The Induction heating system is designed for use under vacuum and low pressure < 0.02 MPa.
  • AttentionA pressure regulator must be installed on your input gas tank ( click picture below to review ) and have the output gas pressure set below 3 PSI at all times after opening the gas tank with the gas maintained in a continuous flowing condition. 
  • Never heat the furnace while the gas release valve is closed and a positive pressure condition exists within the furnace chamber. The pressure gauge must be used to closely monitor the chamber condition at all times during the healing process
  • The flow rate for gases should be limited to < 200 SCCM ( or 200ml/min )